HP710 Infusion Pump
- FOB Price:Get Latest Price >
- Min.Order:4 Set(s)
- Production Capacity:500 pcs/ month
- Payment Terms:T/T
- Favorite
Business Type:Manufacturer
Country/Region:China
Ddu Verified
HOT Rank
8/10
SHENZHEN ELECTECH MEDICAL SUPPLIES CO., LTD.
We are professional supplier of Infusion warmer,Ultrasound gel warmer,Infusion pump,Syringe pump,Infusion Support,Fetal doppler.
Business Type:Manufacturer
Country/Region:China
Ddu Verified
HOT Rank
8/10
Pump: Detachable peristaltic finger pump
Applicable IV set: Exclusive IV Set or standard single PVC IV Set( 10,15,20 and 60 d/mL, with outer diameter 3.4 ~ 4.5mm)
Infusion Mode: ml/h, volume/ time, Bolus, d/min and drug library
Flow rate: 1.0ml/h~1200ml/h
Drip Rate: 1.0d/min~400d/min
Flow rate increment: 1.0~100ml/h with 0.1ml/h increment;100~1200ml/h with 1ml/h increment
~400d/min with 1d/min increment;
VTBI: 0.1~9999.9ml
Accumulated Volume: 0~9999.9ml or>9999.9ml
Time: 1min~>99h59min
Deviation: flow rate: ≤±5%(calibrated IV Set);
Drip rate: ≤±2%
KVO : 0.1~5ml/h adjustable
Purge rate: 600ml/h, displayed infused volume
Occlusion Pressure: Upstream and Downstream pressure detection, range: 40-140Kpa, 1- 10 level selectable with dynamic pressure display.
Air Detector: Ultrasonic air detection, >50μl air bubble, single bubble 50~300 μl, 1-10 level selectable
Volume: 1-10 level selectable
Keypad: Numeric Keypad, user friendly
BOLUS: Bolus Function, flow rate adjustable
Night Mode: LCD become dim in 5min after night mode on.
Self-Testing: Real time self-testing
Alarm: (Audio and Visual) Air in line, upstream occlusion, downstream occlusion, door open, finish, standby time expired, battery low, no battery, battery disconnection, AC Power disconnection, motor error, pump error, up pressure sensor disconnection, down pressure sensor disconnection, parameter error, drop sensor error, drop sensor disconnection, free flow, bottle empty, system abnormal, communication error, IV set not calibrated, air detector error
IV set management and calibration: Open system, acceptable all kinds of IV set after calibration.
Motor Anti- Reverse Function : Motor anti-verse function and monitor the motor constantly.